This is an old revision of the document!
Time Lapse Microscope
Optics
Objectives
Plan Fluor 10x/0.3 (Ph1) 654nm/pixel
Plan Apo 20x/0.75 (DIC N2 / 20X) 327nm/pixel
Plan Fluor 40x/0.75 (DIC N2 / 40X I) 164nm/pixel
Plan Fluor 40x/1.3 Oil (DIC N2/ 40X II) 164nm/pixel
Plan Apo 60x/1.2 WI (DIC N2/ 60X II) 109nm/pixel
Plan Apo 100x/1.4 Oil (Ph3) 65.4nm/pixel
Filter Turret
Semrock DAPI/FITC/TRITC/Cy5 Sedat Quad
GFP/mRFP1
CFP/YFP
FURA2 (Emission Filter within Cube)
D/F/R
Analyzer
Excitation from Lambda 421
LED wavelength | Filter in front of LED |
340 nm | FF01-340/26-25-STR |
380 nm | FF01-389/38-25-STR |
440 nm | FF01-445/40-25-STR |
480 nm | FF01-480/40-25-STR |
530 nm | FF01-520/44-25-STR |
561 nm | FF01-562/40-25-STR |
630 nm | FF01-630/38-25-STR |
Emission Wheel
Fura2 emission filter
Empty
440/11m Semrock Sedat Quad
525/30m Semrock Sedat Quad
607/36m Semrock Sedat Quad
684/24m Semrock Sedat Quad
ET525/50m GFP/mCherry ET (89021 Em2)
ET632/60m GFP/mCherry ET (89021 Em1)
ET470/24m CFP/YFP (89002 Em2)
ET535/30m CFP/YFP (89002 Em1)
Hardware
Nikon Ti-E Microscope
External Phase Unit
Nikon motorized stage
Sutter Lambda 421 LED light source
Sutter Emission wheel (wheel C) controlled by Lambda 10-B (COM5, 128k Baud Rate)
Coolsnap HQ2 (Left port)
Hamamatsu Flash 4.0 (Right port)
In Vivo Scientific incubator and CO2 control
Interface Requirements
10-B XL: COM5 (USB)
10-3: COM2
Firmware
Ti: Combination K
TiControl: 4.32
PC
Software
Excitation Wheel (filters if you used Lambda XL lamp for earlier experiments)
ET340x Fura-2 (79000 Ex1)
ET380x Fura-2 (79000 Ex2)
387/11x Semrock Sedat Quad
485/20x Semrock Sedat Quad
560/25x Semrock Sedat Quad
650/13x Semrock Sedat Quad
ET470/40x GFP/mCherry ET (89021 Ex2)
ET572/35x GFP/mCherry ET (89021 Ex1)
ET430/24x CFP/YFP (89002 Ex1)
ET500/20x CFP/YFP (89002 Ex2)