6D High Throughput Microscope

Photo of 6D/High Throughput scope

Nikon Eclipse Ti-E Motorized Inverted Microscope

Objectives:
Objective Resolution limit (nm) Pixel Size (nm/px) Depth of field (um) Transmitted Light Comments
4x/0.13 air Plan Fluor 2350 1635 76.9 n/a  
10x/0.45 air Plan Apo 678 654 6.42 DIC N1/10x  
20x/0.75 air Plan Apo 407 455 2.31 DIC N2/20x  
40x/0.95 air Plan Apo 321 163.5 1.44 DIC N2/40x I Correction Collar
60x/1.40 oil Plan Apo VC 218 109 0.66 DIC N2/60x II no
100x/1.40 oil Plan Apo VC 218 65.4 0.66 DIC N2/100x I no
2x/0.10 air Plan Apo 3050 3270 130 n/a no
10x/0.30 air Plan Fluor 102 654 14.4 Phase 1 Fura-2
40x/1.30 oil S Fluor 235 163.5 0.77 DIC N2/40x II Fura-2
100x/0.5-1.3 oil S Fluor 610 - 235 65.4 5.2 - 0.77 n/a

Fura-2



Filters:
Fluorophore Excitation Dichroic Emission
DIC halogen 1. Analyzer 2. open
Brightfield halogen 2. Open 2. open
Fura-2 1. D340xv2 3. Fura-2 2. open
Fura-2 2. D380xv2 2. open
CFP 2. ET430/24x 4. CFP/YFP 9. WR470/24m
YFP 1. ET500/20x 10. ET535/30m
GFP 4. S470/30x 5. GFP/mRFP1 7. S510/30m
mRFP1 3. S565/50x 8. S650/75m
DAPI 4. D350/50x 6. Sedat Quad 3. ET455/50m
FITC 3. ET490/20x 4. ET525/36m
Cy3 2. ET555/25x 5. ET605/52m
Cy5 1. ET645/30x 6. ET705/72m


6D_Guide.pdf